Sharp recanalization in the femoral and popliteal arteries
The calcified cap of a chronic total occlusion can be very difficult to cross while staying in the true lumen. Persisting with traditional crossing techniques can create a large subintimal space, exacerbating the initial crossing challenge.
-
Conventional
Wire and support catheter
An .035 Glidewire, angled or straight, or an .014 or .018 wire.
-
Often my go-to
Back end of an .035 Glidewire
Relatively blunt. Cannot be shaped. Tends to stay in the lumen.
-
Still unable to cross
Back end of an .014 or .018 stainless steel wire
Sharp. Can be shaped. Goes outside the lumen more easily.
-
Truly sharp
Chiba needle, 21 gauge, 65 cm
Can be shaped. 1:1 torque. Force and penetration power.
Sharp recanalization, with the back end of a wire or with a needle, is useful when conventional crossing fails. It allows completion of otherwise technically impossible lesions.
A wall of calcium
A solid wall of calcium at the distal superficial femoral artery and proximal popliteal artery can be extremely hard to cross. It often results in an extraluminal crossing, and may need a re-entry device.
In a woman in her 90s with ischemic foot ulcers, CT angiography showed diffuse heavy calcification, with the superficial femoral artery occluded at the adductor canal by a heavily calcified plaque filling the entire lumen of the artery.
Start with a crossing strategy
I start with a crossing strategy: the wire (an .035 Glidewire, angled or straight, or an .014 or .018 wire), a support catheter, antegrade or retrograde access, and an escalation plan. Then I iterate quickly.
In vessels that tend to be straight, like the common femoral, superficial femoral and popliteal arteries, I’ll often start with the back end of an .035 wire.
The back end of an .035 Glidewire
My go-to for crossing is often the back end of an .035 Glidewire. In some instances I don’t even try the front end of a wire. I just go to the back end of a Glidewire, because it tends to stay more luminal than an angled Glidewire or other wires. In my opinion, it is often the best initial crossing wire.
The idea of the back end of an .035 wire is that you don’t have to ram it through. You can probe gently and see what it does. If you feel like you’re staying luminal and making progress, it’s great. If it’s clearly bouncing off calcium and going subintimal, then you’ve got to stop and try something else.
The back end of an .035 Glidewire, regular or stiff, does not create a large subintimal space. It can be supported by a balloon, or run through a 4 French diagnostic catheter with retrograde access as the backup option.
With a balloon
An occluded superficial femoral artery. We were unable to get through using conventional techniques or the back end of the guidewire, because of the wall of calcium. We used the balloon to:
- anchor the catheter
- center the guidewire
Then a very, very brisk push of the wire, and you can see we get through the lesion.
A short popliteal occlusion
In another patient, a short popliteal artery occlusion was crossed with just a gentle probe of the back end of an .035 Glidewire. It crossed in seconds and stayed intraluminal.
Trifurcation reconstruction
In that patient, the trifurcation was then reconstructed with a short, self-expanding nitinol stent in the distal popliteal artery, delivered with its nose cone cut off, and kissing coronary stents into the anterior tibial artery and tibioperoneal trunk. The full case.
The problem with an angled Glidewire
In another patient, with a distal superficial femoral and P1 popliteal occlusion, intravascular ultrasound confirmed a true lumen crossing.
Still unable to cross: the back end of an .014 or .018 wire
Usually I’ll try the back end of an .014 wire rather than going straight to the Chiba needle. It goes outside the lumen more easily, but it does not create a large subintimal space either.
Truly sharp: the Chiba needle
The 21-gauge, 65 cm Chiba needle is the ultimate sharp tool. It crosses almost anything.
It is a simple, cheap and effective tool for crossing difficult occlusions. Staying in the true lumen eliminates the need for re-entry devices, retrograde access and other time-consuming techniques.
In the woman in her 90s with the wall of calcium, crossing catheters and .035 and .014 wires failed. An attempt with the back end of an .014 guidewire was not successful either. A 65 cm 21-gauge Chiba needle was placed through a 40 cm 4 French catheter. Eventually the heavily calcified plaque was crossed. Blood returned from the Chiba needle, and an angiogram confirmed the luminal location. The full case.
Loading, shaping and steering
The Chiba needle comes in 20 cm and 65 cm lengths. You can load it in more than one way:
- Over a wire, an .018 or an .014. You need to be careful when you do that not to shear the tip of the wire off.
- Through an .035 diagnostic catheter: the 65 cm Chiba goes in a 40 cm catheter, or you can cut the hub off a 65 cm catheter. There you can put it in over a wire or on its own, being mindful not to shear the catheter or perforate it.
You can take the needle right out of the box as is, straight, or you can put a subtle curve on it, keeping the trocar in place so you don’t collapse the lumen of the needle. Sometimes you want to shape it and sometimes you don’t: it depends on where the lesion is and how much directional control you need. How much shape you want, and how aggressive you want to be, determines which way you bend it with respect to the bevel face.
The crossing is done under fluoroscopy. You steer the needle by twisting it, and while you are steering you use orthogonal views. Usually the calcium helps you stay within the confines of the artery. Occasionally you’ll perforate the artery, usually without any significance.
Bent along the long axis of the bevel
Very sharp. It wants to engage things.
Bent away from the bevel
Less sharp. Less engaging.
The slight bend, on fluoroscopy
A superficial femoral artery occlusion
A different patient, with an occluded superficial femoral artery.
The crossing, in two steps
A target to aim at
If you can’t figure out how to stay luminal, sometimes placing a target can be really helpful.
A catheter from above
In one trifurcation occlusion, the target was just a catheter from above, in the distal popliteal artery. A Chiba needle from a retrograde posterior tibial artery puncture came up from below to meet it.
A crossing catheter and a looped wire from below
In a thrombosed popliteal artery aneurysm, we were working above and below, and you can see the calcium. A Chiba needle came through a 5 French diagnostic catheter from above, but we couldn’t get true lumen. Then we brought a crossing catheter up from below, and a wire left looped from below became the target. Using both, we were able to target and eventually get through-and-through access and treat the occlusion.
A second needle
For one difficult superficial femoral artery occlusion, we used two Chiba needles: one from above and one from below.
Other uses
In a CERAB reconstruction (covered endovascular reconstruction of the aortic bifurcation) for occlusion of the infrarenal aorta and both common iliac arteries, the aorta could be entered from the left but not from the right. A 20 cm 21-gauge Chiba needle, aimed at an 8 mm balloon inflated from the left, established access from the right.